A Clip-On Logger on a Surgeon's Own Instruments Trained a Policy That Cut in 3 of 4 Live Rabbits. Offline Error Picked the Wrong Designs.
Seoul National University Hospital and Rosota recorded 849 in-vivo demonstrations from hand-held laparoscopic instruments, with no robot in the loop, and deployed the resulting policy with electrosurgery armed. The same paper shows validation error improving while simulated trocar violations climbed from 6% to 69%. For anyone buying demonstration data, that inversion is the finding: accept data on channel timing and closed-loop results, not on how well a model fits it.
A clip-on logger on a surgeon's ordinary laparoscopic instruments recorded 849 usable demonstrations (135 minutes) in four live rabbits, and a diffusion policy trained only on them completed the appendectomy, electrosurgery included, in three of four further rabbits. The robot collected no demonstrations. The paper is arXiv:2609.25625, posted 22 September 2026 by Seoul National University Hospital, Seoul National University and Rosota Inc.
The number that matters more for data work sits in the ablation table. As the authors unfroze more of the vision backbone, validation action error improved at every step (0.152 to 0.142) while simulated trocar violations rose from 6% to 69%. Training the same episodes without per-channel latency matching gave a better validation error, 0.138 against 0.148, and cut the lift stage from 81% to 19%. Offline error ranked the designs backwards.
What was released, and what the logger records
The logger clips onto the shaft outside the patient. One board (Seeed XIAO nRF52840) carries a BNO085 IMU for roll and depression about the trocar and a time-of-flight sensor pointed down the shaft for insertion depth; a WSH137 Hall sensor on the handle reads the jaw. A rigid instrument through a trocar has four degrees of freedom, so these three sensors cover the pose. Against a Franka FR3 driving the instrument, the latency-matched channels track roll and depression to 0.42 and 0.17 degrees RMS and depth to 1.36 mm RMS. A markerless fit of the shaft in the endoscope image on the same rig managed a median 10.7 mm, and the paper reports that a 2.7 mm depth error is already enough to lose the lift stage.
Two corpora were recorded. Ex vivo, on 12 July 2026: rabbit cadaver appendices in a training phantom, two surgeons, 585 episodes recorded and 533 used (74 minutes). In vivo, on 30 August 2026: four live rabbits in one operating-room session, 1,003 episodes recorded, 868 released with complete streams and 849 used after phase-label filters. Both are on Hugging Face under the surgicalrobotics organisation, about 61 GB ex vivo and 46 GB in vivo, with 487 GB of RealSense bags for the simulator. A companion paper, arXiv:2609.25577, describes a camera-fused version of the same logger at 1.21 mm RMS displacement error over 300 trials.
The measurements, with the conditions each was taken under
| Measurement | Value | Condition | Source |
|---|---|---|---|
| In-vivo demonstrations used | 849 of 1,003 recorded (135 min) | 4 live rabbits, one session; 135 dropped for stream quality, 19 for labels | arXiv:2609.25625 Sec. IV-A, IV-C |
| Ex-vivo demonstrations used | 533 of 585 recorded (74 min) | 2 cadavers in a phantom, 2 surgeons | arXiv:2609.25625 Table I; dataset README |
| Appendectomy completed in vivo | 3 of 4 rabbits | shared autonomy, surgeon keyed the phase, energy armed | arXiv:2609.25625 Table III |
| Depth error, latency-matched vs not | 1.36 mm vs 2.69 mm RMS | vs FR3 tip pose, free motion | arXiv:2609.25625 Sec. III-B |
| Lift, with vs without latency matching | 81% vs 19% | 16 simulated episodes each | arXiv:2609.25625 Table II |
| Validation MAE, with vs without latency matching | 0.148 vs 0.138 | same episodes | arXiv:2609.25625 Table II |
| Trocar violations, 0 to 4 unfrozen blocks | 6, 19, 19, 56, 69% | MAE 0.152 falling to 0.142 | arXiv:2609.25625 Table II |
| GR00T-H medical VLA, full approach | 0% (0 of 24 checkpoints closed a jaw on tissue) | same ex-vivo corpus, no phase input | arXiv:2609.25625 Table II, Sec. IV-B |
| Ex-vivo pre-training effect on in-vivo MAE | 1.221 vs 1.413 at 25% data; 1.157 vs 1.173 at 100% | pre-trained vs from scratch | arXiv:2609.25625 Sec. IV-C |
| Vision phase predictor agreement | 24.6% of policy control steps | deployment endoscope | arXiv:2609.25625 Sec. V |
| Dataset licence | "to be stated at release (research use)"; Hugging Face tag license:other | retrieved 24 September 2026 | Rosota-Research/ICRA2027-Invivo-main README |
Why validation error pointed the wrong way
The selection protocol ran 31 candidate designs through 16 closed-loop episodes each in an Isaac Sim reconstruction of the phantom, and scored them on stages (grasp, lift of at least 10 mm, arrive, jaw on tissue) plus one hard failure, the instrument pulling more than 10 mm off the trocar point. Across the unfreezing sweep, validation MAE improved monotonically. Violations rose from 6% to 69%. ResNet-18 had a better MAE than the frozen DINOv3 (0.145 against 0.152) and 88% violations against 6%.
The latency result is the clearest case. Depth lagged attitude by 325 ms on the ex-vivo logger, and the authors attribute the lost lifts to that skew breaking the coordinated insertion lifting needs. Why the unaligned version fits better is not explained in the paper. Our reading, marked as a judgement: in the raw log each label sits beside the smoothed, delayed depth reading that already partly contains it, so it is easier to predict and wrong to execute. Offline error measures agreement with the recording. It does not measure whether the recording describes a motion that can be replayed.
Sixteen episodes is a small sample: the chosen design's 81% has a 95% interval of 57 to 93%, and its advantage over the frozen backbone (81% against 44%) has p = 0.07. The direction of the MAE inversion holds across every axis the authors tried, including capacity and a surgical-domain backbone, so we read it as a pattern rather than a single noisy row.
How much of the operation the policy performed
The headline is accurate and narrow. In each animal the surgeon placed both instruments, checked the hand-over state against the demonstration band, handed over, and supplied the surgical phase by keypress, because the vision phase predictor agreed with the surgeon on only 24.6% of control steps under the deployment endoscope. The policy then ran one rollout from grasp approach to the energised cut: 34.2, 16.8 and 14.7 seconds, 65.7 seconds of policy control in total. In the fourth animal the joint-limit governor stopped the left arm at hand-over and the procedure was finished by hand.
Safety data are good for the portion the policy ran: remote-centre-of-motion error never above 0.84 mm against a 5 mm abort threshold, the cutting jaw closed in all 108 cautery samples, and all four animals alive at one week. Every sample over the threshold that day came under human teleoperation.

What it means for anyone collecting or buying demonstrations
Timing metadata is part of the product. The paper's own ablation shows a corpus losing most of its value (lift 81% to 19%) with no change in content, only in how the channels were aligned, and the loss was invisible to the usual fit metric. A demonstration dataset that ships without per-channel latency, a clock-sync method and absolute timestamps is a dataset the buyer cannot repair. This one ships all three: every stream carries Unix-epoch host time and the in-vivo session files store the clock-fit statistics.
Acceptance tests should be closed-loop. Our judgement: a buyer who scores a delivery by training a model and reading validation error would, on this evidence, have preferred the unaligned version. The authors instead accepted the in-vivo checkpoint through pre-registered event gates (energy false-fire 0.019 against a 0.05 gate, missed-fire 0.032 against 0.30). That is a template worth copying into data contracts.
Capture where the work happens. Ex-vivo pre-training moved in-vivo error only when data were scarce (1.221 against 1.413 at 25% of the corpus) and not at all at full size (1.157 against 1.173), and it did not change one in-vivo gate decision. The authors' conclusion is that live-tissue policies need live-tissue demonstrations. Three of four minimally invasive abdominal operations in a US 2012-2019 inpatient sample were laparoscopic rather than robotic (75.8%), and the paper's point is that all of that instrument motion is currently discarded.
The recorded-but-unused channel is worth noticing. The ex-vivo corpus includes six strain-gauge force channels at 100 Hz (NI6009 files), and the README states the policy does not use them. The in-vivo release lists no force stream. So the only surgical force record in this release is on cadaver tissue, and nobody has yet tested whether it helps.
Two cautions for anyone planning to reuse it. The licence is "to be stated at release (research use)", so treat it as non-commercial until stated otherwise. And the README notes that in-vivo board-to-arm assignments in session.json are wrong for a subset of episodes; the correcting script is promised, not published.
Check it yourself
The ablation table and the latency numbers are in the HTML render. Flatten it and pull the MAE and violation rows:
curl -s https://arxiv.org/html/2609.25625v1 | sed 's/<[^>]*>/ /g' | tr '\n' ' ' | tr -s ' ' | grep -o 'Across the unfreezing sweep.\{0,170\}'
That returns the sentence stating validation MAE improving 0.152 to 0.142 while violations rise from 6% to 69%. Search the same text for 'eliminates successful lifts' for the latency-matching result.
Dataset facts come from the release itself. The README at github.com/Rosota-Research/ICRA2027-Invivo-main gives the episode counts (585 and 1,003 recorded; 868 released in vivo), the file schema and the known issues. Licence tags: curl -s https://huggingface.co/api/datasets/surgicalrobotics/appendectomy-invivo and read cardData.license, which returned other on 24 September 2026. Each in-vivo episode folder holds camera.mp4, camera.h5, one .h5 per logger with imu, tof and hall groups, energy.h5 at 100 Hz, labels.json and session.json.
To reproduce the 65.7 seconds: 34.2 + 16.8 + 14.7 from Table III. To reproduce the 849: 1,003 minus 135 minus 19.
What would prove this wrong
Our reading is that the value of this corpus is carried by its timing alignment and its in-vivo setting, and that offline fit is an unreliable acceptance test for demonstration data of this kind. Two tests would break that.
First, with a date: by 30 September 2027, if an independent group trains on the released in-vivo corpus and reports that the checkpoint with the lowest validation error is also the one with the fewest closed-loop failures, across at least three candidate designs, then the inversion here was specific to 16-episode simulation noise rather than a property of the data.
Second: if someone deploys a policy trained on video-estimated instrument kinematics (the SurgiPose route, 6 to 12 mm error) and matches three of four completed in-vivo procedures, the instrument-mounted logger is not what made this work, and the case for paying for instrumented capture in surgery weakens.
Sources
- Yee, Oh, Lee et al., From Instrument-Mounted Demonstrations to In-Vivo Execution: Learning Bimanual Laparoscopic Appendectomy Without Robot-Collected Demonstrations, arXiv:2609.25625, 22 September 2026
- arXiv HTML render: Sections III-B, IV-B to IV-D, Tables I to III
- Lee, Yee, Oh, Lee et al., Recording Hand-Held Laparoscopic Instrument Motion in the Operating Room, arXiv:2609.25577, 22 September 2026
- Dataset release README, Rosota-Research/ICRA2027-Invivo-main (licence: to be stated at release), retrieved 24 September 2026
- In-vivo corpus on Hugging Face, surgicalrobotics/appendectomy-invivo, license:other
- Ex-vivo corpus on Hugging Face, surgicalrobotics/appendectomy-exvivo, license:other
- Nelson, Chen, Haworth et al., Open-H-Embodiment (source of the GR00T-H baseline), arXiv:2604.21017
- Chi, Xu, Pan et al., Universal Manipulation Interface (the latency-budget rule the paper extends), RSS 2024
- BLO LAB, a surgeon dataset's GR00T fine-tune lost to 'predict no motion' in all 11 held-out arms
Building or training robots?
We license manipulation datasets and run custom capture programs. Get in touch to see what fits.

A Closed-Form Weight Edit Took pi0.5 From 27 to 106 of 128 on a Misread Instruction, but Only When Written Into the VLM

V-JEPA 2's Predictor Adds 0.04 Top-1 Over Copying the Present on 27,000 Egocentric Trials. By Horizon, It Runs From Minus 0.05 at 1 s to Plus 0.16 at 3.5 s.