A Clip-On Logger on a Surgeon's Own Instruments Trained a Policy That Cut in 3 of 4 Live Rabbits. Offline Error Picked the Wrong Designs.

Seoul National University Hospital and Rosota recorded 849 in-vivo demonstrations from hand-held laparoscopic instruments, with no robot in the loop, and deployed the resulting policy with electrosurgery armed. The same paper shows validation error improving while simulated trocar violations climbed from 6% to 69%. For anyone buying demonstration data, that inversion is the finding: accept data on channel timing and closed-loop results, not on how well a model fits it.

A clip-on logger on a surgeon's ordinary laparoscopic instruments recorded 849 usable demonstrations (135 minutes) in four live rabbits, and a diffusion policy trained only on them completed the appendectomy, electrosurgery included, in three of four further rabbits. The robot collected no demonstrations. The paper is arXiv:2609.25625, posted 22 September 2026 by Seoul National University Hospital, Seoul National University and Rosota Inc.

The number that matters more for data work sits in the ablation table. As the authors unfroze more of the vision backbone, validation action error improved at every step (0.152 to 0.142) while simulated trocar violations rose from 6% to 69%. Training the same episodes without per-channel latency matching gave a better validation error, 0.138 against 0.148, and cut the lift stage from 81% to 19%. Offline error ranked the designs backwards.

What was released, and what the logger records

The logger clips onto the shaft outside the patient. One board (Seeed XIAO nRF52840) carries a BNO085 IMU for roll and depression about the trocar and a time-of-flight sensor pointed down the shaft for insertion depth; a WSH137 Hall sensor on the handle reads the jaw. A rigid instrument through a trocar has four degrees of freedom, so these three sensors cover the pose. Against a Franka FR3 driving the instrument, the latency-matched channels track roll and depression to 0.42 and 0.17 degrees RMS and depth to 1.36 mm RMS. A markerless fit of the shaft in the endoscope image on the same rig managed a median 10.7 mm, and the paper reports that a 2.7 mm depth error is already enough to lose the lift stage.

Two corpora were recorded. Ex vivo, on 12 July 2026: rabbit cadaver appendices in a training phantom, two surgeons, 585 episodes recorded and 533 used (74 minutes). In vivo, on 30 August 2026: four live rabbits in one operating-room session, 1,003 episodes recorded, 868 released with complete streams and 849 used after phase-label filters. Both are on Hugging Face under the surgicalrobotics organisation, about 61 GB ex vivo and 46 GB in vivo, with 487 GB of RealSense bags for the simulator. A companion paper, arXiv:2609.25577, describes a camera-fused version of the same logger at 1.21 mm RMS displacement error over 300 trials.

From the surgeon's instrument to a training pair: four channels, four different delays ON THE INSTRUMENT MEASURED DELAY (ex-vivo logger, vs FR3) TRAINING PAIR BNO085 IMU shaft roll + depression 24-26 ms, pure transport ToF rangefinder insertion depth d = l - r 349 ms, 93% first-order lag WSH137 Hall sensor jaw opening, 200 Hz 79 ms, firmware loop Endoscope video policy image input 42 ms, by cross-correlation rewind each channel by its own delay, resample on image clock 10 Hz, 16-step chunks Depth lags attitude by 325 ms, 3.5 policy steps at 10 Hz. No single time shift removes that skew. Per-channel rewinding cut dynamic depth error from 2.69 mm to 1.36 mm against the FR3 tip pose. In simulation, the same 533 episodes without it: lift 19% instead of 81%, and a LOWER validation error. In-vivo logger (BLE, battery): depth 57-63 ms, attitude 3-9 ms; the endoscope's 55 ms became the largest term.
The capture chain. Each channel reaches the log late by a different amount, and the gap between depth and attitude is larger than a policy step.

The measurements, with the conditions each was taken under

MeasurementValueConditionSource
In-vivo demonstrations used849 of 1,003 recorded (135 min)4 live rabbits, one session; 135 dropped for stream quality, 19 for labelsarXiv:2609.25625 Sec. IV-A, IV-C
Ex-vivo demonstrations used533 of 585 recorded (74 min)2 cadavers in a phantom, 2 surgeonsarXiv:2609.25625 Table I; dataset README
Appendectomy completed in vivo3 of 4 rabbitsshared autonomy, surgeon keyed the phase, energy armedarXiv:2609.25625 Table III
Depth error, latency-matched vs not1.36 mm vs 2.69 mm RMSvs FR3 tip pose, free motionarXiv:2609.25625 Sec. III-B
Lift, with vs without latency matching81% vs 19%16 simulated episodes eacharXiv:2609.25625 Table II
Validation MAE, with vs without latency matching0.148 vs 0.138same episodesarXiv:2609.25625 Table II
Trocar violations, 0 to 4 unfrozen blocks6, 19, 19, 56, 69%MAE 0.152 falling to 0.142arXiv:2609.25625 Table II
GR00T-H medical VLA, full approach0% (0 of 24 checkpoints closed a jaw on tissue)same ex-vivo corpus, no phase inputarXiv:2609.25625 Table II, Sec. IV-B
Ex-vivo pre-training effect on in-vivo MAE1.221 vs 1.413 at 25% data; 1.157 vs 1.173 at 100%pre-trained vs from scratcharXiv:2609.25625 Sec. IV-C
Vision phase predictor agreement24.6% of policy control stepsdeployment endoscopearXiv:2609.25625 Sec. V
Dataset licence"to be stated at release (research use)"; Hugging Face tag license:otherretrieved 24 September 2026Rosota-Research/ICRA2027-Invivo-main README
All values are from the paper, its dataset README or the Hugging Face API. Nothing here was re-measured by us.

Why validation error pointed the wrong way

The selection protocol ran 31 candidate designs through 16 closed-loop episodes each in an Isaac Sim reconstruction of the phantom, and scored them on stages (grasp, lift of at least 10 mm, arrive, jaw on tissue) plus one hard failure, the instrument pulling more than 10 mm off the trocar point. Across the unfreezing sweep, validation MAE improved monotonically. Violations rose from 6% to 69%. ResNet-18 had a better MAE than the frozen DINOv3 (0.145 against 0.152) and 88% violations against 6%.

The latency result is the clearest case. Depth lagged attitude by 325 ms on the ex-vivo logger, and the authors attribute the lost lifts to that skew breaking the coordinated insertion lifting needs. Why the unaligned version fits better is not explained in the paper. Our reading, marked as a judgement: in the raw log each label sits beside the smoothed, delayed depth reading that already partly contains it, so it is easier to predict and wrong to execute. Offline error measures agreement with the recording. It does not measure whether the recording describes a motion that can be replayed.

Sixteen episodes is a small sample: the chosen design's 81% has a 95% interval of 57 to 93%, and its advantage over the frozen backbone (81% against 44%) has p = 0.07. The direction of the MAE inversion holds across every axis the authors tried, including capacity and a surgical-domain backbone, so we read it as a pattern rather than a single noisy row.

Unfreezing more DINOv3 blocks: validation error fell every step, trocar violations rose 0 20 40 60 80 100 share of 16 simulated episodes (%) 6% 44% 0 blocks MAE 0.152 19% 81% 1 block MAE 0.148 19% 50% 2 blocks MAE 0.146 56% 38% 3 blocks MAE 0.143 69% 31% 4 blocks MAE 0.142 trocar violation full approach, no violation Latency matching, same data with: lift 81%, MAE 0.148 without: lift 19%, MAE 0.138 Encoder, full approach ResNet-18: 12%, MAE 0.145 GR00T-H VLA: 0%, MAE 0.184 DINOv3, 1 block: 81%, MAE 0.148 The lowest-error option lost in every sweep.
Table II of the paper as a chart. Every step that lowered validation error made the simulated policy less safe.

How much of the operation the policy performed

The headline is accurate and narrow. In each animal the surgeon placed both instruments, checked the hand-over state against the demonstration band, handed over, and supplied the surgical phase by keypress, because the vision phase predictor agreed with the surgeon on only 24.6% of control steps under the deployment endoscope. The policy then ran one rollout from grasp approach to the energised cut: 34.2, 16.8 and 14.7 seconds, 65.7 seconds of policy control in total. In the fourth animal the joint-limit governor stopped the left arm at hand-over and the procedure was finished by hand.

Safety data are good for the portion the policy ran: remote-centre-of-motion error never above 0.84 mm against a 5 mm abort threshold, the cutting jaw closed in all 108 cautery samples, and all four animals alive at one week. Every sample over the threshold that day came under human teleoperation.

Abstract technical rendering of four thin signal traces running at slightly different phase offsets along a rigid metallic shaft, converging to a single aligned bundle at one end.
Illustration, not data. Four sensor streams with different delays, aligned into one training sample; the measured delays are in the first diagram.
Deployment on four live rabbits: what the policy did and what the surgeon did Appendectomy completed Policy control time Cautery fired by policy Peak RCM error Hand-over channels in demo band Rabbit 1 yes 34.2 s 4.6 s 0.20 mm 6/6 Rabbit 2 yes 16.8 s 3.0 s 0.21 mm 4/6 Rabbit 3 yes 14.7 s 3.2 s 0.84 mm 4/6 Rabbit 4 no, aborted not reached not reached not reached not reported Surgeon, every animal placed instruments, checked hand-over, keyed the phase Policy, 3 animals 65.7 s total, grasp approach to energised cut, 5 Hz Phase predictor agreed with the surgeon on 24.6% of control steps Rabbit 4: the joint-limit governor stopped the left arm before any policy command reached tissue. Completed by hand.
Table III of the paper. The policy performed the grasp-to-cut segment; the surgeon owned placement and phase in every animal.

What it means for anyone collecting or buying demonstrations

Timing metadata is part of the product. The paper's own ablation shows a corpus losing most of its value (lift 81% to 19%) with no change in content, only in how the channels were aligned, and the loss was invisible to the usual fit metric. A demonstration dataset that ships without per-channel latency, a clock-sync method and absolute timestamps is a dataset the buyer cannot repair. This one ships all three: every stream carries Unix-epoch host time and the in-vivo session files store the clock-fit statistics.

Acceptance tests should be closed-loop. Our judgement: a buyer who scores a delivery by training a model and reading validation error would, on this evidence, have preferred the unaligned version. The authors instead accepted the in-vivo checkpoint through pre-registered event gates (energy false-fire 0.019 against a 0.05 gate, missed-fire 0.032 against 0.30). That is a template worth copying into data contracts.

Capture where the work happens. Ex-vivo pre-training moved in-vivo error only when data were scarce (1.221 against 1.413 at 25% of the corpus) and not at all at full size (1.157 against 1.173), and it did not change one in-vivo gate decision. The authors' conclusion is that live-tissue policies need live-tissue demonstrations. Three of four minimally invasive abdominal operations in a US 2012-2019 inpatient sample were laparoscopic rather than robotic (75.8%), and the paper's point is that all of that instrument motion is currently discarded.

The recorded-but-unused channel is worth noticing. The ex-vivo corpus includes six strain-gauge force channels at 100 Hz (NI6009 files), and the README states the policy does not use them. The in-vivo release lists no force stream. So the only surgical force record in this release is on cadaver tissue, and nobody has yet tested whether it helps.

Two cautions for anyone planning to reuse it. The licence is "to be stated at release (research use)", so treat it as non-commercial until stated otherwise. And the README notes that in-vivo board-to-arm assignments in session.json are wrong for a subset of episodes; the correcting script is promised, not published.

Check it yourself

The ablation table and the latency numbers are in the HTML render. Flatten it and pull the MAE and violation rows:

curl -s https://arxiv.org/html/2609.25625v1 | sed 's/<[^>]*>/ /g' | tr '\n' ' ' | tr -s ' ' | grep -o 'Across the unfreezing sweep.\{0,170\}'

That returns the sentence stating validation MAE improving 0.152 to 0.142 while violations rise from 6% to 69%. Search the same text for 'eliminates successful lifts' for the latency-matching result.

Dataset facts come from the release itself. The README at github.com/Rosota-Research/ICRA2027-Invivo-main gives the episode counts (585 and 1,003 recorded; 868 released in vivo), the file schema and the known issues. Licence tags: curl -s https://huggingface.co/api/datasets/surgicalrobotics/appendectomy-invivo and read cardData.license, which returned other on 24 September 2026. Each in-vivo episode folder holds camera.mp4, camera.h5, one .h5 per logger with imu, tof and hall groups, energy.h5 at 100 Hz, labels.json and session.json.

To reproduce the 65.7 seconds: 34.2 + 16.8 + 14.7 from Table III. To reproduce the 849: 1,003 minus 135 minus 19.

What would prove this wrong

Our reading is that the value of this corpus is carried by its timing alignment and its in-vivo setting, and that offline fit is an unreliable acceptance test for demonstration data of this kind. Two tests would break that.

First, with a date: by 30 September 2027, if an independent group trains on the released in-vivo corpus and reports that the checkpoint with the lowest validation error is also the one with the fewest closed-loop failures, across at least three candidate designs, then the inversion here was specific to 16-episode simulation noise rather than a property of the data.

Second: if someone deploys a policy trained on video-estimated instrument kinematics (the SurgiPose route, 6 to 12 mm error) and matches three of four completed in-vivo procedures, the instrument-mounted logger is not what made this work, and the case for paying for instrumented capture in surgery weakens.